|
|
The ECG is a result of potentials appearing at the surface of the chest
as a result of the summation of depolarisation and repolarisation of many
myocardial fibres simultaneously.
|
|
Depolarisation initiated at the SA node spreads as a wavefront across
the two atria, and also at a higher speed along the three inter-nodal tracts to
the AV node. There is a delay at the AV node, then the wavefront travels at
high speed down the His bundle in the interventricular septum, dividing the two
ventricles. The His bundle branches into the Purkinje system which conducts the
wavefronts along much of the endocardial surfaces of the two ventricles. The
wavefronts then spread more slowly through the normal myocardium. They spread
from the inside to the outside of the ventricles (Figure 83). Wavefronts travel
with higher velocity in the direction of the fibre orientation. The morphology
of the resulting ECG recorded on the chest surface depends on the orientation
of the heart, the active recording electrode and the reference electrode
(Figure 84).
The signal waveform produced for each heart beat consists of the P
wave, due to atrial depolarisation, the QRS complex due to ventricular
depolarisation, and the T wave caused by ventricular repolarisation. The
effects of summing all the electrical activity in the heart can be represented
by an electrical dipole whose magnitude and direction is constantly changing
(Figure 85). The scalar magnitude of the ECG is then the dot product of the
dipole and the electrode orientation.
|
|
Generally 12 lead positions are commonly used to record the ECG (Figure
86). The first three are known as leads I, II and III. They are left arm to
right arm for I, ie the active lead is on the left arm (usually the left wrist)
and the reference electrode is on the right arm (wrist). Lead II is left leg
(ankle) to right arm, and lead III is left leg to left arm. The right ankle is
usually grounded. The lead vectors can be represented by an equilateral
triangle, known as Eindhoven's triangle. The direction of lead I vector is 0
degrees by convention. The direction of lead II is 60 degrees and lead III is
120 degrees.
|
|
|
|
The remaining lead positions use a common reference, known as Wilson's
central terminal (Figure 87). This consists of the right and left wrists joined
to the left ankle, each through a suitably large resistor, eg 5 M ohms. The
first three active electrodes are the right and left wrists and the left ankle.
In practice this means that when one limb is an active electrode, it is shunted
by the resistance that is part of the Wilson's central terminal circuit. To
avoid this shunting, the active limb is connected by a resistor of half the
value of the others, to the non-inverting input of the amplifier. The limb is
not connected to the Wilson's central terminal. This is known as the augmented
lead system (Figure 88). The leads are called aVL, aVR and aVF for active lead
connections to the left wrist, right wrist and left ankle respectively. The
other six leads also use the Wilson's central terminal as reference and the
active lead is placed in different positions over the front of the chest
overlying the base of the heart and the apex.
|
|
Electrocardiograph recorders require a number of features to make them
usable in clinical practice. These are:
1. Protection circuitry against defibrillation shocks that may be given
to the patient. These shocks may be up to 3,000 volts.
2. Lead selector. The default mode is to automatically record all 12
leads simultaneously. Otherwise one or more leads are selected for recording.
In cheaper machines, three or four leads are recorded simultaneously for five
seconds at a time, with automatic switching to each group of three or four
leads.
3. Calibration signal of 1 mV is automatically applied to each channel
for a brief period.
4. Preamplifiers have a very high input impedance and high common mode
rejection ratio (reject signals appearing on both the active and reference
leads simultaneously).
5. Isolation circuit separates the patient from the power supply.
6. Driven right leg circuit. In older instruments the right leg was
grounded. Now, as part of the isolation, the right leg is not connected to
ground, but is instead driven by an amplifier to remain at a virtual ground.
7. Driver amplifier follows the pre-amplifier and drives the chart
recorder. It also filters the signal to remove any dc offset and high frequency
noise.
8. Microprocessor system contains circuitry for digitising the signal,
and storing and analysing it. Most systems can automatically calculate the
rate, analyse most of the common arrhythmias, report the axes of some features,
and detect old and recent myocardial infarcts (heart attacks, coronary
occlusions).
9. Recorder printer is used to provide a hard copy of the ECG, together
with the patient information and the analysis and diagnosis.
Digitising and filtering is done so as to prevent aliasing, that is the
false representation of a high frequency signal as a low frequency one (Figure
89). This happens if the sampling and digitising rate of the signal is less
than twice the highest frequency in the signal, as determined by the filter. In
that case, signals of higher frequency are digitised as low frequency signals.
|
|
|
|
Frequency distortion can occur if the filters are not set so that the
pass band ranges from 0.02 Hz to 150 Hz (Figure 90). If there is too much dc
offset on the input signal, the output of the amplifiers may saturate, clipping
the signal (Figure 91). Ground loops may induce 50 Hz noise on the recording if
other equipment is connected to the patient but has a separate earthing system
(Figure 92).
High voltage shocks or similar transients can cause the amplifiers to
saturate temporarily, producing artefacts on the recording. Noise may be
capacitively or inductively coupled to the recording leads (Figure 93).
Electrode movement can also cause noise. All electrodes can be
represented by a battery in series with a parallel resistor and capacitor. The
battery represents the polarisation voltage produced by dissimilar active and
reference electrode materials being in contact with an electrolyte, the saline
solution of body fluid. Alternatively the polarisation voltage can be seen as
due to the layer of ions produced at the electrode by the current flowing
between the electrode and the body fluid. The resistance is due to the
resistance between the electrode and the tissue of interest. The capacitance is
formed by the metal electrode plate as one electrode, deeper tissue as the
second electrode and body fluid as the electrolyte.
The polarisation voltage is an unwanted dc offset that can swamp the
signal being recorded. Any op amp dc bias current flowing through through the
patient's body will also cause a dc voltage drop across the resistance of the
electrode. In addition this dc current can charge the equivalent capacitor,
also producing an unwanted dc offset.
|
|
Electrodes used to record ECG's consist of metal disks with a
conductive gel between the disk and the skin to ensure good contact. The metal
disk and the skin form two plates of a capacitor and the conductive gel is the
electrolyte. If the capacitor is charged and the distance between the disk and
skin changes due to movement, the capacitance changes, so for a given charge
the voltage must change. This produces movement noise. This can be overcome by
making the electrode of an inverted cup made of insulating material with the
metal electrode at the top. The inside is filled with conductive gel. The
insulated edge of the cup sits on the skin so that the distance between the
electrode and the skin remains constant. This prevents the changes in
capacitance which cause movement noise.
|
|
There are two main consequences of coronary heart disease: arrhythmia
and myocardial infarction. An arrhythmia is an abnormal rhythm developing in
the heart. This may be either a slow rhythm or a fast rhythm. A myocardial
infarction is a localised region of dead muscle due to a loss of blood supply.
|
|